Operation
Multi 3430
45
ba75865e19
04/2015
*
Brand names or trade names are trademarks of their respective
owners protected by law.
Note
The buffers are selected in the menu, pH /
<MENU/ENTER>
/
Calibra-
tion
/
Buffer
(see section 4.7.3).
Calibration points
Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The meter determines the
following values and calculates the calibration line as follows:
Note
You can display the slope in the units, mV/pH or % (see section 4.7.3).
Stability control
The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be termi-
nated at any time (accepting the current value).
Calibration record
The new calibration values are displayed when a calibration procedure
is completed.
23
Chemsolute
*
2.000
4.000
7.000
10.000
20 °C
24
USABlueBook
*
4.000
7.000
10.000
25 °C
25
YSI
*
4.000
7.000
10.000
25 °C
No.
Buffer set*
pH values
at
Determined
values
Displayed calibration data
1-point
Asy
Zero point =
Asy
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asy
Slp.
Zero point =
Asy
Slope =
Slp.
3-point to
5-point
Asy
Slp.
Zero point =
Asy
Slope =
Slp.
The calibration line is calculated
by linear regression.
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......