Operation
Multi 3430
113
ba75865e19
04/2015
The sensor settings are reset under the
Reset
menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
<MENU/ENTER>
key.
Conductivity
The following settings for conductivity measurements are reset to the
default settings with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
<MENU/ENTER>
key.
Salinity (function)
Off
Setting
Default settings
Setting
Default settings
Cal. interval
150 d
Measured parameter
ϰ
Cell constant (c)
Depending on the connected
measuring cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set up)
0.100 cm
-1
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of the
linear temperature compensation
2.000 %/K
TDS factor
1,00
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......