Carrying out an automatic calibration (AutoCal)
Make sure that in the sensor menu,
Buffer
menu, the buffer set is cor-
rectly selected (see section 4.7.3).
Use any one to five buffer solutions of the selected buffer set in ascend-
ing or descending order.
Below, calibration with Technical buffers (
TEC
) is described. When
other buffer sets are used, other nominal buffer values are displayed.
Apart from that, the procedure is identical.
Note
If single-point calibration was set in the menu, the calibration procedure
is automatically finished with the measurement of buffer solution 1 and
the calibration record is displayed.
1
In the measured value display, select the measured parameter
pH or mV with
<M>
.
2
Start the calibration with
<CAL>
.
The calibration display for the first buffer appears (voltage dis-
play).
3
Thoroughly rinse the IDS sensor with deionized water.
4
Immerse the IDS pH sensor in buffer solution 1.
5
For measurements without temperature sensor
(e.g. when using an IDS adapter
)
:
Measure the temperature of the buffer manually and enter it with
<
><
>
.
6
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
15.03.2014 08:00
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......