Multi 3430
What to do if...
118
ba75865e19
04/2015
Data transmission to
USB memory device
does not work
Data transmission to
USB printer device does
not work
Error message,
Only use rechargeable
batteries!
Ni-MH 1.2 V, >2100 mAh
CAUTION
Unsuitable batterie types can cause damage during the charging
process.
Cause
Remedy
– Connected USB memory device
has not been recognized, e. g.
– the USB memory device has
been formattet to a file sys-
tem which is not supported
(e. g. NTFS)
– Use other USB memory
device
– Reformat the USB memory
device to FAT 16 or FAT 32
file system
(Caution: Reformatting
erases all data on the USB
memory device. Back up all
data before reformatting.)
Cause
Remedy
– The USB-B interface is con-
nected to a PC
– Disconnect the PC from the
USB-B interface
– Connected USB memory device
does not work
– Use a suitable USB printer
– Check printer settings (see
– No power pack connected
– Connect the power pack
Cause
Remedy
– A voltage that is unusual for
NiMH rechargeable batteries
was found.
– Disconnect the power pack
from the meter.
– Check whether there are
suitable rechargeable batter-
ies (
Ni-MH 1.2 V, >2100
mAh
) in the battery compart-
ment.
–
Unsuitable batteries
:
Insert suitable recharge-
able batteries and recon-
nect the power pack.
–
Suitable batteries
:
Connect the power pack
and close the error mes-
sage with
<[ESC]>
or
<M>
.
Содержание WTW Multi 3430
Страница 1: ...OPERATING MANUAL ba75865e19 04 2015 Multi 3430 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3430 6 ba75865y19 04 2015 9 2 Firmware Update for IDS Sensors 132...
Страница 14: ...Multi 3430 Overview 14 ba75865e19 04 2015...
Страница 18: ...Multi 3430 Safety 18 ba75865e19 04 2015...
Страница 22: ...Multi 3430 Commissioning 22 ba75865e19 04 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 5...
Страница 124: ...Multi 3430 What to do if 124 ba75865e19 04 2015...
Страница 133: ......