7.5.3
Controller information
7.5.4
cal-test
The
cal-test
function is used to test the sealing of the system measuring
head - rubber sleeve - sample bottle and the operability of the system
OxiTop
®
Control.
In the
BOD Standard
mode with the
GLP
On
setting, the controller indi-
cates when the next cal test is due, depending on the set calibration
interval (see section 7.4.4 GLP (O
page 45).
To perform the test, you need the WTW test resource, OxiTop
®
PM,
order number 209 333.
Sample preparation: see operating manual of the OxiTop
®
PM test
resource.
CHECK
menu.
Use
<
><
>
to select the
-
Controller Info
menu item.
The display shows the following
information:
Number of measuring heads
for which there is still memory
capacity
Status of the supply batteries
Status of the data backup
batteries (
OK
/
LOBAT
).
Use
<
><
>
to scroll to the
specification of the
Software version
Series number
(can differ from the example
shown here)
Return to the
CHECK
menu:
With
<RUN/ENTER>
.
RUN/ENTER
CHECK
menu.
Use
<
><
>
to select the
- Cal
Test
menu item.