GLP/TOOLS
OxiTop
®
Control OC 100
56
ba31114e07
08/2014
The user interface of the controller guides you through the test:
7.6
Maintenance
CHECK
menu.
Use
<
><
>
to select the
-
Pneumatic test
menu item.
Set the plunger of the syringe on
the OxiTop
®
PT test resource to
5 scale parts.
Tightly screw the measuring
head to be tested onto the Oxi-
Top
®
PT test resource.
(Exceeding the time causes a
return to the
CHECK
menu).
(Exceeding the time causes a
return to the
CHECK
menu).
The controller display shows the
result of the pneumatic test.
Use
<RUN/ENTER>
to return to
the
CHECK
menu.
RUN/ENTER
RUN/ENTER
RUN/ENTER
GLP/TOOLS
menu.
Use
<
><
>
to select the
-
Maintenance
menu item.