78
ba31114e07
08/2014
What to do if...
OxiTop
®
Control OC 100
The IR printer does not
react
The printer is not switched on
The printer has no optical contact
with the controller:
– Distance too great, angular posi-
tion imprecise
– IR window of the printer averted
from the controller
– Shading by other objects.
Printer batteries are empty.
No paper or the wrong paper is
loaded.
Printing is not possible in the oper-
ating state of the controller
selected.
The printer or the controller is
defective.
Switch on the printer
Establish optical contact
Check or change the printer batter-
ies.
Please read the operating instruc-
tions of the printer.
Check in the operating manual of
the controller whether printing is
possible in the state selected.
Note: During printing, the message
Printing active ...
always appears
on the display
Cause
Remedy