Operating mode: BOD Standard / BOD Routine
OxiTop
®
Control OC 100
22
ba31114e07
08/2014
Confirm the selected measure-
ment range for the sample. The
automatically assigned sample
number (YY/MM/DD and
sequential number) is given in
the header line.
Additional information:
Type of measurement, run duration, measurement range,
final date, ID number.
Temp
display => the
AutoTemp
, page 41) is switched on.
GLP
display => the
GLP
TOOLS, page 41) is switched on.
Change the ID number for the additional identification of the
sample (e.g. sampling location) as follows:
Select the
- I.D. number
menu item with
<
><
>
Confirm
with
<RUN/ENTER>
.
Set the required Id number (setting range 001 ... 255) with
<
><
>
.
Confirm with
<RUN/ENTER>
.
Use
<PRINT>
to print out the entire sample information
(in
BOD Routine
mode only), see section 6 P
Confirm the start of measure-
ment. Contact selection:
The controller repeatedly sends the start information in the
scanner mode until feedback is successfully received from the
OxiTop
®
measuring head. After the successful start message
from the measuring head:
Displays
! started !
From this
point in time, the sample exists
within the
SAMPLE STATUS
TYPE
list (
sample manage-
ment).
Subsequently, the request to
start the next measuring head is
made automatically.
In
BOD Routine
operating mode there is no
1... identifier in
the display, and the
controller automatically returns to the entry menu after starting
the measuring head.
RUN/ENTER
RUN/ENTER
Содержание OxiTop Control OC 100
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 100 6 ba31114e07 08 2014...
Страница 14: ...Commissioning OxiTop Control OC 100 14 ba31114e07 08 2014...
Страница 20: ...Operation OxiTop Control OC 100 20 ba31114e07 08 2014...
Страница 82: ...OxiTop Control OC 100 82 ba31114d07 08 2014...
Страница 83: ......