Technical data
OxiTop
®
Control OC 100
80
ba31114e07
08/2014
11.2
Measuring head OxiTop
®
-C or OxiTop
®
-C/B
Measuring principle
Piezoresistive pressure sensor
Operation pressure
range
500 ... 1350 hPa (mbar).
Maximum allowed pres-
sure
2000 hPa (mbar)
Accuracy
± 1% of the measured value ± 1 hPa
Resolution
1 hPa
Display
LED pilot lamp
Power supply
2 x lithium battery CR2430
Battery runtime
Typically 2 years, in case of normal use
Protective class
3, instrument safety: EN 61010-1
Type of protection
IP 54
EMC
EN 61326-1, FCC Class A
Ambient temperature
Storage: -25°C ... +65°C
Operation: +5°C ... +50°C
Dimensions
H: 70 mm,
Ø
70 mm
Weight
95 g
Test certificates
CE
Содержание OxiTop Control OC 100