OxiTop
®
Control OC 100
Operating mode: BOD Standard / BOD Routine
ba31114e07
08/2014
33
5.5
Evaluation
- Missing
:
Only appears when the run duration of the sample has
expired. Here you can remove the data of a missing or defec-
tive measuring head from the
SAMPLE STATUS TYPE
list
(
sample management) of the controller. In the
BOD Standard
operating mode, you can set a sample to the "finished" status
(checkmark ) if all the other measuring heads are already “fin-
ished”: Erases the missing measuring heads from the
SAMPLE STATUS TYPE
list (
sample management).
SAMPLE STATUS TYPE
list
(sample management)
Use
<
><
>
to select a sam-
ple.
Use
<PRINT>
to print out the
results with curves.
1 End of measuring range
2 Mean value
Evaluation of the sample:
Display of all the curves together
with mean value data.
Check the display for outliers.
Use
<PRINT>
to print the results
with curves (according to print-
ing format - see
the section 6 P
Example: Overall sample with
outlier.
In the
BOD Routine
operating mode, the evaluation shows the
selected sample as a single curve with the measured value
data.
Use
<RUN/ENTER>
to switch between the cursor interroga-
tion (see section 5.5.3 C
, page 36) dis-
play of the curve with measured value data.
Scroll through the parallel sam-
ples: Display of individual curves
(cyclical pass) with data of the
BOD final value.
Use
<PRINT>
to print out a sin-
gle result with a single curve.
1
2
Содержание OxiTop Control OC 100
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 100 6 ba31114e07 08 2014...
Страница 14: ...Commissioning OxiTop Control OC 100 14 ba31114e07 08 2014...
Страница 20: ...Operation OxiTop Control OC 100 20 ba31114e07 08 2014...
Страница 82: ...OxiTop Control OC 100 82 ba31114d07 08 2014...
Страница 83: ......