OxiTop
®
Control OC 100
What to do if...
ba31114e07
08/2014
73
0 of 19
called up
(sample message, no. is
variable)
No measuring head was found by
the controller.
The measuring heads are placed
at another location (e.g. other lev-
els in the incubator, other incuba-
tors, further storage locations)
The measuring heads have no
optical contact with the controller.
– Distance too great, angular
position imprecise
– IR window of the measuring
head averted from the control-
ler
– Optical shading by other
objects
The controller is defective.
Modification of the distances or
angle
Search at other storage locations
Check the controller (see section
10.3 G
Display message
Cause
Remedy