OxiTop
®
Control OC 100
Operating mode: BOD Standard / BOD Routine
ba31114e07
08/2014
35
5.5.2
Excluding a curve (
BOD Standard
operating mode)
This function is used to temporarily exclude a single curve (e.g. an
outlier) from the evaluation and averaging of an overall sample.
The function "List of measuring heads“ (see section 5.4
SAMPLE STATUS TYPE
), page 26) is used
to find leaky or defective single samples.
The curve is only excluded temporarily!
The data stock of the
SAMPLE STATUS TYPE
list (
sample management) does not
change. The excluded curve is present again when the call is
repeated.
Display of single curves:
Select curve.
Use
<
><
>
to select
-
Exclude curve
.
- Exclude curve
is preselected.
- Back
: Returns to the previous
menu.
Message
Curve
excluded!
.
Updated display (curve
excluded, mean value recalcu-
lated).
Use
<PRINT>
to print the
updated results with curves
(without the excluded curve).
RUN/ENTER
RUN/ENTER
RUN/ENTER