Calling up all data
This function is used to call up the data of all measuring heads, no
matter in which operating mode they were started.
To call up the data of individual measuring heads: see section 5.4
SAMPLE STATUS TYPE
SELECT ACTION
menu (com-
munication with the measuring
heads):
Preselected:
- Start sample
.
Select the
- Call up all data
menu
item with
<
><
>
Point the controller at the measuring
heads:
Queries the data of all active
measuring heads in the scanner
mode.
The controller stores the data
obtained and updates the
SAMPLE STATUS TYPE
list
(
sample management).
Duration of a pass: approx. 3
seconds for 12 measuring heads
in a stirring system.
40 cm to 1 m
RUN/ENTER
If the measuring heads do not all respond in the 1st pass of the data query,
the controller searches for the missing measuring heads in the scanner mode
for approximately a further 7 seconds. Without having to press another key,
you can continue the data queries at other locations (e.g. other stirring plat-
forms or other thermal cabinets).
Approximately 7 seconds after the last call up was performed, the scanner
mode breaks off automatically and the following display appears:
Содержание OxiTop Control OC 100
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 100 6 ba31114e07 08 2014...
Страница 14: ...Commissioning OxiTop Control OC 100 14 ba31114e07 08 2014...
Страница 20: ...Operation OxiTop Control OC 100 20 ba31114e07 08 2014...
Страница 82: ...OxiTop Control OC 100 82 ba31114d07 08 2014...
Страница 83: ......