OxiTop
®
Control OC 100
GLP/TOOLS
ba31114e07
08/2014
45
7.4.3
Date/time
Here you set the date and time for the controller (important for the
assignment of sample numbers).
7.4.4
GLP
(Operating mode
BOD Standard
)
The GLP (Good Laboratory Practice) set of rules requires a detailed
description of all equippings and procedures in a laboratory.
These descriptions include the room equipment of the laboratory, the
instruments used, and the measuring procedures laid down, and
require the complete documentation of the samples tested. The type
and frequency of the use of test resources is to be described and docu-
mented as well.
To make the documentation of the test resource monitoring easier and
simpler the OxiTop
®
Control system has a
GLP
mode.
When this mode is switched on, the following data are documented:
the calibration intervals that have been set
the calibrations that have been performed
The reports contain the series numbers of the measuring head and
controller and the date of the calibration.
The
GLP
mode monitors the calibration intervals and blocks the start of
a new measurement if a calibration will be due within the measuring
time.
SETTINGS
menu:
Use
<
><
>
to select the
-
date/time
menu item.
Use
<
><
>
to set each num-
ber block on a black back-
ground.
Confirm with
<RUN/ENTER>
.
Set the day, month, year, min-
utes, seconds one after the
other.
RUN/ENTER
The run duration of samples already started is not affected by
changing the date or time.
Содержание OxiTop Control OC 100
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 100 6 ba31114e07 08 2014...
Страница 14: ...Commissioning OxiTop Control OC 100 14 ba31114e07 08 2014...
Страница 20: ...Operation OxiTop Control OC 100 20 ba31114e07 08 2014...
Страница 82: ...OxiTop Control OC 100 82 ba31114d07 08 2014...
Страница 83: ......