background image

MICRO-ePUMP

World Precision Instruments 

19

The following table is useful for converting between different units of volume.

Cubic Measure

1 cm

3

1 mm

3

100 µm

3

10 µm

3

1 µm

3

Volume

1 mL

1 µL

1 nL

1 pL (10

-12 

L) 1 fL (10

-15 

L)

For ejected volumes less than 1 nL, visual inspection in air proves to be difficult due 

to rapid evaporation. The same technique may be used though if the drop is kept 

submerged under oil. Droplets may seem to disappear after emergence from the tip. 

Sometimes this is due to creepage of the aqueous fluid back along the outside shank of 

the micropipette. This creepage may be decreased by silanizing the outside of the pipette. 

TIP

: Precise assays of ejected volume may also be obtained through various 

radioisotopic methods. See the bibliography for further information.

A slight deflection of the micropipette tip may be noticed during the application of the 

pressure pulse. This deflection may be eliminated by ensuring that the micropipette is 

firmly seated in the holder and that the holder is firmly held by a manipulator.

Calibrating Rate using a Known Volume

We can calculate the volume of liquid contained in a 1 mm length of a glass pipette, if 

we know the inner diameter (ID). Then, we can inject that known volume and measure 

the time it takes to inject it. From this we can calculate the rate of injection. For 

example, if the volume in 1 mm of glass with 0.58 µm ID (WPI #

1B100

) is 264 nL, and 

it takes 30 seconds to inject that volume, then the flow for 1 second is 264 nL/30 s or 

8.8 nL/second, and a 1 nL volume then takes 0.113 seconds to inject.
This method is discussed in the following JoVE video: http://www.jove.com/video/2079/

intravenous-microinjections-zebrafish-larvae-to-study-acute-kidney.

In the Jove video, the capillary tip is 10–20 µm. By following these steps, we are able to 

calculate the timing needed to deliver a 1 nl injection by counting the time between 1 

mm marks as the fluid is injected. 

NOTE

: The 

MICRO-ePUMP

 timing should not be adjusted to go below 10–15 ms as an 

absolute minimum, since it takes 6–10 ms for the pressure valves to respond.
1.  Fill a small dish with mineral oil and place it under a stereo microscope.
2.  Turn on your air and vacuum pumps, and set your compensation pressure and 

your injection pressure. The proper settings must be determined experimentally. 

Turn on the 

MICRO-ePUMP

.

 NOTE

: The injection pressure must exceed the compensation pressure in order 

to eject the fluid.

3.  Use a pipetter to inject a 10 µl sphere of fluid into the mineral oil.
4.  Using a fine pointed permanent marker, mark the injecting micropipette at 1 mm 

increments.

5.  Mount your “graduated” micropipette on the 

MICRO-ePUMP

 that is connected to 

the Injection Port on the 

MICRO-ePUMP

.

6.  Set the mode to Manual. Hold down the 

foot switch to inject fluid from the 

micropipette into the oil-filled dish.

Содержание MICRO-ePUMP

Страница 1: ...WORLD PRECISION INSTRUMENTS MICRO ePUMP Microinjector Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 012220...

Страница 2: ......

Страница 3: ...mpensation Pressure 16 Manufacturing Micropipettes 18 Calibrating Volume by Measuring Droplet 18 Calibrating Rate using a Known Volume 19 Multibarrel Microinjection 20 MAINTENANCE 21 Cleaning 21 ACCES...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...ION Designed to simplify intracellular injection and a variety of other micro in jec tion tasks the MICRO ePUMP uses carefully regulated air pres sure for injecting cells with fluid Injected volumes r...

Страница 6: ...nce being injected A reference electrode is used to place the media at 0 0 V potential with reference to the generated voltage The MICRO ePUMP offers separate regulated compensation back filling preve...

Страница 7: ...r Adapter 1 Instruction Manual available online at www wpiinc com manuals Unpacking Upon receipt of this instrument make a thorough inspection of the contents and check for possible damage Missing car...

Страница 8: ...pette tip a parameter to modify it using the knob or the settings area to adjust the settings Adjustment Knob Rotate the adjustment knob to change the highlighted parameter Press it to toggle between...

Страница 9: ...switch icon Press the foot switch to initiate a timed sequence timed mode or to manually eject pressure manual mode This foot switch is a two stage switch Depress it half way down to engage the MICRO...

Страница 10: ...used protocols in the MICRO ePUMP memory If the box is empty no protocol is loaded Injection Counter Every time you depress the foot switch all the way when the port is enabled the injection counter...

Страница 11: ...ration with minimal trauma The researcher determines the amplitude and frequency of the signal that best suits the application If you are using the MICRO ePORE techcnology touch the parameter and rota...

Страница 12: ...eters as desired Then press one of the Save buttons For example press SAVE1 to save the parameters as Protocol 1 To load a set of saved parameters a protocol press the appropriate LOAD button The numb...

Страница 13: ...ng connected to the barb Body Fig 11 The 300744 Capillary Kit must be properly assembled 1 Slide the bare end of the Adapter Tubing Assembly through the Pipette Handle 2 Connect it to the barb of the...

Страница 14: ...with a slight clockwise twist Fig 12 Left Align the connector with the Injection Port slide the connector in position and rotate the connector a little clockwise to lock it into place Fig 13 Right Th...

Страница 15: ...electrode Fig 17 For the probe style electrode the black pin of the microelectrode interface cable connects to the back side of the reference electrode block The probe style electrode pin plugs into...

Страница 16: ...ameter by turning the knob To change the dial adjustment speed press the knob on the front of the MICRO ePUMP The snail changes the display in 0 01 increments and the running man changes it in increme...

Страница 17: ...this configuration No parameters show in the bottom half of the screen 2 Tap a parameter and turn the knob to change the parameter Fig 24 Touch the Frequency parameter or the Amplitude parameter and...

Страница 18: ...shows the injection count Whenever the foot switch is fully depressed the injection counter increments Reset it to zero by pressing on the Injections area of the main display touchscreen Adjust the M...

Страница 19: ...from the main screen press the SAVE1 SAVE2 or SAVE3 button The parameters are saved in Protocol 1 Protocol 2 or Protocol 3 depending on which SAVE button you chose 4 Press the BACK button to return to...

Страница 20: ...Capillary Kit on page 9 5 When the micropipette is properly installed click on the Settings area to access the Settings menu and press the Port OFF button to enable the Injection Port TECHNIQUES IN MI...

Страница 21: ...the contact angle for glass and water is zero unless the glass surface is treated From this equation we see that the smaller the inside diameter the greater the capillary action The capillary pressure...

Страница 22: ...on becomes prohibitively large and the hydrophilic surface of the glass greatly limits the flow of fluid through the tip Silanization decreases the surface tension and allows the fluid to flow smoothl...

Страница 23: ...te of injection For example if the volume in 1 mm of glass with 0 58 m ID WPI 1B100 is 264 nL and it takes 30 seconds to inject that volume then the flow for 1 second is 264 nL 30 s or 8 8 nL second a...

Страница 24: ...vital to making a correct injection of 1 nL 11 Divide the average found in step 9 by the calculated volume in nL of a 1 mm length of your pipette found in step 10 This establishes the time in millise...

Страница 25: ...nd other contamination There is no need to clean the inside Use a soft cloth dampened with a mild solution of detergent and water Do not use abrasive cleaners ACCESSORIES Part Number Description 75122...

Страница 26: ...n Port connector call WPI technical support 104 Tank pressure is too low to deliver the desired injection pressure Tank depleted due to previous injections or leak Wait for pump to charge Check all pr...

Страница 27: ...ping Weight 90 264 V 50 60 Hz 48 W 213 w x 207 D x 250 H mm 8 4 x 8 1 x 9 8 5 kg 11 lb Ambience Ambient Temperature Relative Humidity For indoor use only 15 C 40 C 10 75 non condensing Both Backfillin...

Страница 28: ...1 mm 1 24 mm 100 nL 464 m 575 36 m 10 nL 215 m 267 m 1 nL 100 m 1 24 m 100 pL 46 m 57 m 10 pL 21 5 m 27 m 1 pL 10 m 12 4 m Fig 37 This graph compares the dimensions of a cube and a sphere with the sam...

Страница 29: ...ing of ion sensitive microelectrodes J Neuroscience Methods 8 231 247 1983 Deyhimi F and J A Coles Rapid silylation of a glass surface Choice of reagent and effect of experimental parameters on hydrop...

Страница 30: ...26 World Precision Instruments DECLARATION OF CONFORMITY...

Страница 31: ...special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or...

Страница 32: ...n Germany Tel 49 0 6031 67708 0 Fax 49 0 6031 67708 80 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 27A Liangfeng Building No 8 Dongfang Road Pudong Free Trade Zone Shang...

Отзывы: