background image

MICRO-ePUMP

World Precision Instruments 

17

dye is dissolved in the injection fluid. The capillary effect may be observed with the 

color change at the tip of the pipette. When the fluid flows into the pipette, the color 

of the tip becomes lighter. If the compensation pressure is higher than the capillary 

pressure, the fluid oozes out of the pipette. The solution around the pipette will be 

colored. Adjusting the compensation pressure prevents this from happening. 
Understanding how 

capillary action causes the front filling of the pipette helps you to 

correctly use the compensation pressure. The flow rate is determined by the capillary 

action and the tip size. Since the tip size is often determined by the requirement of 

the application, controlling the compensation pressure becomes the main option to 

eliminate the uptake of fluid by capillary action. 

NOTE:

 

The pressure of capillary action is determined by the inner diameter of 

the glass capillary where the meniscus of air/liquid interface is located. It has 

nothing to do with the pipette tip size.

If you assume the pipette tip is a cylindrical shape, the pressure of capillary action can 

be described by the LaPlace equation:

P=4

γ

 cos

θ

/d

γ

 = surface tension 

θ

 = contact angle between the water and glass 

d

 = inner diameter of the capillary where the meniscus is

 

located

In most cases, we can assume the contact angle for glass and water is zero (unless 

the glass surface is treated). From this equation we see that the smaller the inside 

diameter, the greater the capillary action. The capillary pressure can vary a thousand 

times when the meniscus is moved from a 0.5 µm ID tip to the 0.5 mm shank. 

The pressure at 0.5 µm tip is about 80 PSI (in aqueous solution) while at the shank will 

be only 0.08 PSI. Using one regulator to counterbalance the pressure in such a large 

dynamic range is not practical. 10 PSI can counterbalance a meniscus at the section of 

tip where the inner diameter is 4 µm. 
In practice, this is the highest pressure ever needed. On the lower pressure end, it 

becomes difficult to exactly counterbalance the capillary pressure when the meniscus 

is at the shank of the pipette. However, a 0.1–0.2 PSI pressure imbalance will not 

cause a significant problem if the tip is small enough. The gravity of the fluid and the 

flow resistance caused by friction from the glass wall will both help to stop the solution 

flow at this pressure level. 

TIP

: If the lowest pressure setting is ineffective at preventing the pipette fluid from 

leaking out, try setting the compensation pressure to zero to see if the gravity and 

friction are sufficient to counterbalance the front fill. 

TIP

: The capillary action can also be reduced by adding a hydrophobic fluid (such as 

silicone oil) behind the hydrophilic saline solution. It can be completely eliminated by 

silanizing the shank of the pipette (silanization increases the 

θ

 in the LaPlace equation 

to 90º).

Содержание MICRO-ePUMP

Страница 1: ...WORLD PRECISION INSTRUMENTS MICRO ePUMP Microinjector Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 012220...

Страница 2: ......

Страница 3: ...mpensation Pressure 16 Manufacturing Micropipettes 18 Calibrating Volume by Measuring Droplet 18 Calibrating Rate using a Known Volume 19 Multibarrel Microinjection 20 MAINTENANCE 21 Cleaning 21 ACCES...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...ION Designed to simplify intracellular injection and a variety of other micro in jec tion tasks the MICRO ePUMP uses carefully regulated air pres sure for injecting cells with fluid Injected volumes r...

Страница 6: ...nce being injected A reference electrode is used to place the media at 0 0 V potential with reference to the generated voltage The MICRO ePUMP offers separate regulated compensation back filling preve...

Страница 7: ...r Adapter 1 Instruction Manual available online at www wpiinc com manuals Unpacking Upon receipt of this instrument make a thorough inspection of the contents and check for possible damage Missing car...

Страница 8: ...pette tip a parameter to modify it using the knob or the settings area to adjust the settings Adjustment Knob Rotate the adjustment knob to change the highlighted parameter Press it to toggle between...

Страница 9: ...switch icon Press the foot switch to initiate a timed sequence timed mode or to manually eject pressure manual mode This foot switch is a two stage switch Depress it half way down to engage the MICRO...

Страница 10: ...used protocols in the MICRO ePUMP memory If the box is empty no protocol is loaded Injection Counter Every time you depress the foot switch all the way when the port is enabled the injection counter...

Страница 11: ...ration with minimal trauma The researcher determines the amplitude and frequency of the signal that best suits the application If you are using the MICRO ePORE techcnology touch the parameter and rota...

Страница 12: ...eters as desired Then press one of the Save buttons For example press SAVE1 to save the parameters as Protocol 1 To load a set of saved parameters a protocol press the appropriate LOAD button The numb...

Страница 13: ...ng connected to the barb Body Fig 11 The 300744 Capillary Kit must be properly assembled 1 Slide the bare end of the Adapter Tubing Assembly through the Pipette Handle 2 Connect it to the barb of the...

Страница 14: ...with a slight clockwise twist Fig 12 Left Align the connector with the Injection Port slide the connector in position and rotate the connector a little clockwise to lock it into place Fig 13 Right Th...

Страница 15: ...electrode Fig 17 For the probe style electrode the black pin of the microelectrode interface cable connects to the back side of the reference electrode block The probe style electrode pin plugs into...

Страница 16: ...ameter by turning the knob To change the dial adjustment speed press the knob on the front of the MICRO ePUMP The snail changes the display in 0 01 increments and the running man changes it in increme...

Страница 17: ...this configuration No parameters show in the bottom half of the screen 2 Tap a parameter and turn the knob to change the parameter Fig 24 Touch the Frequency parameter or the Amplitude parameter and...

Страница 18: ...shows the injection count Whenever the foot switch is fully depressed the injection counter increments Reset it to zero by pressing on the Injections area of the main display touchscreen Adjust the M...

Страница 19: ...from the main screen press the SAVE1 SAVE2 or SAVE3 button The parameters are saved in Protocol 1 Protocol 2 or Protocol 3 depending on which SAVE button you chose 4 Press the BACK button to return to...

Страница 20: ...Capillary Kit on page 9 5 When the micropipette is properly installed click on the Settings area to access the Settings menu and press the Port OFF button to enable the Injection Port TECHNIQUES IN MI...

Страница 21: ...the contact angle for glass and water is zero unless the glass surface is treated From this equation we see that the smaller the inside diameter the greater the capillary action The capillary pressure...

Страница 22: ...on becomes prohibitively large and the hydrophilic surface of the glass greatly limits the flow of fluid through the tip Silanization decreases the surface tension and allows the fluid to flow smoothl...

Страница 23: ...te of injection For example if the volume in 1 mm of glass with 0 58 m ID WPI 1B100 is 264 nL and it takes 30 seconds to inject that volume then the flow for 1 second is 264 nL 30 s or 8 8 nL second a...

Страница 24: ...vital to making a correct injection of 1 nL 11 Divide the average found in step 9 by the calculated volume in nL of a 1 mm length of your pipette found in step 10 This establishes the time in millise...

Страница 25: ...nd other contamination There is no need to clean the inside Use a soft cloth dampened with a mild solution of detergent and water Do not use abrasive cleaners ACCESSORIES Part Number Description 75122...

Страница 26: ...n Port connector call WPI technical support 104 Tank pressure is too low to deliver the desired injection pressure Tank depleted due to previous injections or leak Wait for pump to charge Check all pr...

Страница 27: ...ping Weight 90 264 V 50 60 Hz 48 W 213 w x 207 D x 250 H mm 8 4 x 8 1 x 9 8 5 kg 11 lb Ambience Ambient Temperature Relative Humidity For indoor use only 15 C 40 C 10 75 non condensing Both Backfillin...

Страница 28: ...1 mm 1 24 mm 100 nL 464 m 575 36 m 10 nL 215 m 267 m 1 nL 100 m 1 24 m 100 pL 46 m 57 m 10 pL 21 5 m 27 m 1 pL 10 m 12 4 m Fig 37 This graph compares the dimensions of a cube and a sphere with the sam...

Страница 29: ...ing of ion sensitive microelectrodes J Neuroscience Methods 8 231 247 1983 Deyhimi F and J A Coles Rapid silylation of a glass surface Choice of reagent and effect of experimental parameters on hydrop...

Страница 30: ...26 World Precision Instruments DECLARATION OF CONFORMITY...

Страница 31: ...special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or...

Страница 32: ...n Germany Tel 49 0 6031 67708 0 Fax 49 0 6031 67708 80 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 27A Liangfeng Building No 8 Dongfang Road Pudong Free Trade Zone Shang...

Отзывы: