22
World Precision Instruments
press the yellow banner to re-enable the port.
On power up the
MICRO-ePUMP
performs a diagnostic self-test. If there is a failure
during the self-test, an error code displays and the unit stops the sequence. It will not
progress into the operational mode unless it is powered off and then on, and the self-
test completes without error.
ERROR Condition
Possible Cause
Action
101
A leak was detected
while the instrument
was idle.
Faulty pressure
connection or broken
pipette or tubing.
Check all pressure connections
and make sure the pipette is not
broken.
102
Low pumping pressure
was detected.
Faulty pressure
connection or broken
pipette or tubing.
Check all pressure connections
and make sure the pipette is not
broken.
Pump failure.
If the problem persists with
nothing connected to the
Injection Port connector, call
WPI technical support.
103
Pumping time to fully
charged pressure
exceeded 5 minutes.
Faulty pressure
connection or broken
pipette or tubing.
Check all pressure connections
and make sure the pipette is not
broken.
Pump failure.
If the problem persists with
nothing connected to the
Injection Port connector, call
WPI technical support.
104
Tank pressure is too
low to deliver the
desired injection
pressure.
Tank depleted due to
previous injections or
leak.
Wait for pump to charge. Check
all pressure connections and
make sure the pipette is not
broken.
200
Drain valve self-test
failed
Drain Valve of
connections defective.
Call WPI technical support.
201
Failed to charge tank
during self-test at low
pressure.
Charging time
exceeded.
Ensure nothing is connected
to the Injection Port. Call WPI
technical support if the problem
persists.
202
Failed to charge to the
maximum pressure
during the self-test.
Charging time
exceeded.
Ensure nothing is connected to
Injection Port. Call WPI technical
support if the problem persists.
203
High setting port valve
test failed during the
self-test.
Failed to reach the
proper Injection Port
pressure.
Ensure nothing is connected to
Injection Port. Call WPI technical
support if problem persists.
204
Low setting port valve
test failed during the
self-test.
Failed to reach the
proper Injection Port
pressure.
Ensure nothing is connected to
Injection Port. Call WPI technical
support if problem persists.
NOTE
: If you have a problem/issue with that falls outside the definitions of this
troubleshooting section, contact the WPI Technical Support team at (941) 371-1003 or
Содержание MICRO-ePUMP