MICRO-ePUMP
World Precision Instruments
9
Fig. 8—(Left) The power up screen shows the software revision on your MICRO-ePUMP.
Fig. 9—As the self-test runs through the programmed diagnostic sequence, the display
shows the results.
3. Press the touchscreen to access the main screen. By default, the injection port is
disabled. Tap the warning or the Settings icon to enable the Injection Port.
Fig. 10—The warning indicates that the injection port is disabled by default on startup.
Assembling the Capillary Kit
Pipette Holder
Pipette Handle
Tubing Assembly
Adapter Tubing Assembly
Gasket
Cap
Pipette
Tubing connected to the barb
Body
Fig. 11—The 300744 Capillary Kit must be properly assembled.
1. Slide the bare end of the Adapter Tubing Assembly through the Pipette Handle.
2. Connect it to the barb of the Pipette Holder Body. Be careful so that the barb
does not break when attaching or removing the tubing.
3. Slide the Pipette Handle over the barb and carefully screw it onto the Body of the
Pipette Holder. Be careful not to cross thread it.
4. Place a gasket of the correct size in the cap. Refer to the table below. Then, insert
the blunt end of the micropipette (pulled capillary glass) into the cap through
the gasket and into the body. Screw the cap in place. The screw cap and rubber
gasket firmly hold the glass micropipette.
Gasket Color
Green Black
Red
White
Pipette Diameter (mm)
1.0
1.2
1.5
1.65