52
6. Disassemble two screws and replace a
new motor as the figure shows.
7.9
Error Code 0206
Motor Speed Fault: Data transmission problems, asynchronous communication and
transmission between Main I and Main II board are not synchronised.
There is a connection problem between the centrifugal motor and the Main II board.
1. Push out the drawer slowly with both
hands. Avoid breaking the driving gear of
motor, as shown in the figure.
2. Unscrew the screws on the bottom of the
drawer and take out the switch board.
3. Check the FPC connecting status.
Confirm it connects properly.