49
5)
If the above steps are invalid, replace the location module directly.
1. Take out the whole optical module and
remove the location LED.
2.
Unscrew the screw on the location
detector as shown on the figure.
3.
Then, take out the location module and
replace it with a new one.
7.7
Error Code 0109
Optical Output is Unstable: The ambient temperature is too low or high, being beyond the
range of temperature that the analyser can operate properly. This causes irregular
temperature changes in the cavity of the lamp which leads to unstable output of the light
source.
The xenon lamp output source may be unstable due to transformer or circuit component
changes, etc. This fault is very rare, however if this problem occurs, please contact Woodley
Equipment Company.
7.8
Error Code 0202 & 0203
Error Code 0202 Drawer Open Fault: After 5 seconds of sending out drawer open
command, the disc lock has been released but the drawer is still in the chamber.