44
5. Use side cutting pliers to cut the ribbons of the heating film and the temperature sensor,
as shown in the figure.
6. Unplug the five connectors shown in the red circle in the figure.
7. Unplug the three connectors in the red circles as shown in the figure.
8.
After taking out the temperature sensor in the red circle, push out the drawer slowly
with both hands. Avoid breaking the driving gear on the motor.