Touchscreen is Not Responding
Touch on the touchscreen and other places react. It may be due to the compression
deformation of the sponge isolation pad of the LCD screen, resulting in the touch point on
the touchscreen contacting with the injection parts or the failure of the touchscreen. Please
reset the touch screen after restarting the device. If you still have not solved this problem,
please contact Woodley Equipment Company.
8.6
System Automatically Restarts
The device is automatically restarted repeatedly.
8.6.1
Reasons
1)
Insufficient or unstable power supply.
2)
Kernel program error.
3)
Power strip failure.
8.6.2
Troubleshooting
Please follow the steps in order. After each step is finished, retest whether the fault is
cleared or not. If not, proceed to the next step.
1)
Check whether the power plug is plugged in tightly.
2)
Replace the adaptor to test for adaptor failure.
3)
Unplug the power cord between the internal power supply board and the main board of the
equipment, check the metal contact in each plug and re-plug.
4)
Replace the core board.
5)
Replace the power board.
8.7
System Breakdown
The interface has no response.
8.7.1
Reasons
1)
Within 3 minutes of the test countdown, the system's computational complexity is large
and the time required for exiting is too long
2)
After clicking the button, the debugging interface needs to wait for some time to enter.