59
4. When the speed of the motor is stable,
press the
Location
function and get 30
data of any inclined planes. Check the
thirty data. It is normal that the bigger
one (Zero-point) is near 650. If the bigger
one is near other data (near
400) that means the disc is leaking and
you will need retest with a new disc.
5. If the bigger data and others decrease
together and the Zero-point is less than
550, clean the spacers at the front of the
LED. Then increase the IR value. Press
the IR Setting and then Save to
make the change of the IR active.
3)
Check the FPGA cable and plug connections. If the problem cannot be solved, replace the
it with a new one.
4)
Check the centrifugal motor.
1.
Enter the debugging function of the
analyser and choose the
Motor Control
interface.
2.
Press
Open
and let the drawer open. Put
in a disc and close the drawer.
3.
Click
Motor Startup
to rotate with
1200RPM.
4. Select
Next Page
and choose 20 times of
the
Forward Start
and
Backward Start
switching to test the centrifugal motor at
a high speed.