Section 11 Reagent Disc Error and Abnormal Results
Result shows [---]
The test result is incorrect or a negative value.
Reasons:
1)
The blood sample is abnormal and does not enter the
reagent cuvette.
2)
The intensity of the analyser detection wavelength is
abnormal
LDL - C reports [---]
The test result is incorrect or a negative value.
Reasons:
1)
An abnormal blood sample results in a TG greater than
4.52 and the result of LDL-C cannot be calculated.
2)
The blood sample is abnormal and does not fill the reagent
cuvettes.
3)
The intensity of the analyser detection wavelength is
abnormal
Na
+
reports [---]
-
The test result is incorrect or a negative value.
Reasons:
1)
An abnormal blood sample results in an incorrect result.
2)
The intensity of the analyser detection wavelength is
abnormal
Cl
-
reports [---]
-
The test result is incorrect or a negative value.
Reasons:
1)
The reagent wells of two identical reagents in the disc
have large deviations
2)
An abnormal blood sample results in an incorrect result.
3)
The intensity of the analyser detection wavelength is
abnormal
Results appear >=
The test item detection exceeds the upper limit.
Reasons:
1)
An abnormal blood sample results in an incorrect result.
2)
The intensity of the analyser detection wavelength is
abnormal