34
8. After updating the factor press
Save
,
and then
Saved Successfully
will
appear.
9. If a wrong number is input or you
make an error and want to change
back to the original factors you had
before, you need to input the
reciprocal of the ratio.
For example
,
the TP given a wrong
coefficient of 2, you can input 0.5
(the reciprocal of 2) as a new
factor to change, and press
Save
.
Notice
: After you make a save, all the coefficient will change to 1, as a basic number for
the next time.
And if you want input a negative number, you can press the dot for more than 2 seconds
on the keyboard.
This action affects the results of the entire batch of reagents, so be sure to perform all
tests with appropriate quality controls.