69
2.
Remove the Main I board. Unscrew two
screws that fix the core card. Replace with
a new core board, assemble back to the
Main I board and check if the status of the
core board is normal. The core board must
be installed in the right direction.
5)
Check the drive board of the LCD panel and whether the connection line is normal.
1.
Exchange another upper panel for a test
if the problem can be solved. If it is valid,
then change a new panel directly.
6)
Replace the power supply board, refer to section 7.17.
8.2
White Screen (BIOS Cannot Start)
The screen displays white after startup.
8.2.1
Reasons
Program error.
8.2.2
Troubleshooting
Please follow the steps in order. After each step is finished, retest whether the fault is cleared or
not. If not, proceed to the next step.
Replace the core board. Copy the “config.bin” file of the received device parameters to the root
directory of the USB flash disc. Plug in the USB flash disc to restart the device.