15
Backward Start
to change the
direction of the motor’s rotation.
Switch the direction of rotation 20
times. During this process, if the
analyser automatically shuts down or
restarts due to excessive motor
current, a new motor is needed.
10.LED Test
Press
Next Page
to access the LED
Test function.
If increasing the IR value in step 7
(maximum 255) cannot adjust the
maximum value in 30 data near 650
±
30, diagnosis of the LED should be
made with this function. Touch
LED
Test
and automatically acquire data.
If the difference between the two
numbers is more than 500, the LED
has problems and it needs to be
replaced.
3)
The default
function is the function of clearing all test records and analyser operation
records.
1. Default Function Introduction
The other hidden functions in the
Settings is the
Default
function.