Error Code 0203 Drawer Close Fault: After 5 seconds of sending the drawer close
command, the drawer still hasn’t closed.
7.8.1
Reasons for Error 0202 and 0203
1)
The position switch on the location control wire is in the wrong state, may be damaged or
the assembly has caused the mechanical switch to not move smoothly.
2)
The drawer control wire (with motor) is loose at the signal plug, resulting in a false signal
connection.
3)
Drawer drive gear is damaged, resulting in abnormal transmission.
7.8.2
Troubleshooting
Please follow the steps in order. After each step is finished, retest whether the fault is
cleared or not. If not, proceed to the next step.
1.
Disassemble the machine enclosure and check whether all the cables are properly
plugged in, focusing on the drawer location wire and the drawer drive motor wire.