72
3)
Too much memory.
4)
Incorrect positioning of touchscreen.
5)
LCD panel fault.
6)
Missing core programs.
8.7.2
Troubleshooting
Please follow the steps in order. After each step is finished, retest whether the fault is
cleared or not. If not, proceed to the next step.
1)
Restart the device and observe.
2)
Check the touchscreen and recalibrate.
3)
Replace the LCD panel drive board and recalibrate it.
4)
Replace the upper panel. (Woodley Equipment provides the assembled upper panel with the
LCD screen connected to line).
5)
Replace the core board. Copy the “config.bin” file of the received device parameters to the
root directory of the USB flash disc. Plug in the USB flash disc to restart the device.
8.8
System Does Not Start Properly
The equipment cannot start up normally after being powered on. Check whether the power
cord plug is plugged in tightly. Reinsert the power cord again. Note that the power plug
should be inserted into the root of the connector of the device and restart the device.
8.9
System Returns to the Main Interface During Detection Process
After starting detection, the system returns to the Main Interface and the test fails.
8.9.1
Reasons
1)
Centrifugal motor does not start.
2)
The disc is not flattened and is not stuck to the bottom.
8.9.2
Troubleshooting
1)
Upgrade the hardware to the latest version.
2)
Clean the windows at front of the LED.
3)
Adjust the IR value, confirm whether the zero value is normal during debugging.