![WLD-TEC Fuego SCS Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/wld-tec/fuego-scs/fuego-scs_instruction-manual_990733052.webp)
52
4.1 Cooling time reminder
The cooling time reminder is helpful to determine
the exact cooling down time of e.g.: inoculation
loops. In the background, this function counts
down a pre-set cooling down time. Once the
cooling down time has expired, a short acoustic
signal sounds and a reminder symbol appears
on the display.
To set the cooling down time in the reminder
functions menu, select the inoculation loop symbol
and confirm by a push on the function knob (1).
In the following menu item the cooling down time
can be set between 1 and 25 seconds. If "-- sec"
is selected, the function is deactivated and no reminder
will be given (factory default setting: "-- sec").
Another push on the function knob (1) memorizes
the time set and the menu item is exited.
The acoustic signal can be deactivated in the
settings menu (see paragraph 5.5).
i
Fig. 4.1/1
Adjusted cooling down time
Reminder symbol
Cooling time reminder
4.2 Gas consumption display
Using this function, a gas duration specific to one
cartridge type can be set. In this case, during
operation a "Gas cartridge" symbol appears which
empties according to the burn time. Once the time
has expired, a reminder is given to notify the user
to have a new cartridge at hand. The gas consump-
tion display is independent of the user accounts.
The settings apply to both user accounts.
This reminder is an INDICATION that the
cartridge in use could be empty soon.
This reminder is NOT A DEMAND to
change the cartridge immediately!! Note
in particular that gas cartridges without
a built-in valve should only be changed
if the gas cartridge is completely empty!
When changing the cartridge, ensure
adequate ventilation.
For setting the gas duration in the reminder functions menu select the "Gas cartridge"
symbol and confirm with a push on the function knob (1). Now in the 1st step the kind
of cartridge can be selected by turning the function knob (1). Depending on the cartridge
selected, previously specified times taken from a database appear on the display.
Confirm the type of cartridge by pressing the function knob (1). In the 2nd step the bur-
ning time can be individually adjusted to achieve even more precise timing if necessary.
By pressing the function knob (1) the value set is memorized and the menu item is exited.
1. Step: Choose cartridge
2. Step: Adjust time if necessary
Gas consumption display
Fig. 4.2/1
!