DPD 1 A and DPD 1 B reagent MUST be added to the vial BEFORE water sample is added to
avoid misreadings!
Turbidity caused by high concentration of calcium ions will affect the measurement result. To
prevent that, please use DPD HC (High Calcium) reagents.
Liquid reagents should be stored at temperatures of 5 - 10 °C in securely closed bottles.
Shake the liquid reagent thoroughly before adding the liquid to the vial.
Do not clean glassware and equipment with household detergents, as these could greatly
reduce the measurement result. To prevent any contamination, the cuvette, lid and stirrer
should be stored in a 0.1 % sodium hypochlorite solution for one hour. Then rinse with distilled
water, thoroughly.
The DPD reagent buffers the pH value of the sample water in the range of 6.2 to 6.5, which is
essential for colour development. If the sample water is very alkaline or acidic, this must be
adjusted to a pH within the range of 6 to 7 by addition of 0.5 mol/l sulphuric acid or 1 mol/l
caustic soda, respectively, before the DPD reagent is added.
Make sure no chlorine escapes while preparing the sample. The measurement must be
performed directly after sampling.
If the sample water contains further oxidizing agents, these will react like chlorine and contribute
to the measurement result.
If the sample water contains more than 20 mg/l chlorine, a measured value of 0 mg/l can be
displayed. In that case, a dilution is recommended.
T28/204
T28/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...