(147)
Legionella
(Cou Test)
60 - 1000000
cfu/test (Leg)
Reagent-Kit
Internal Name: 147-Legionella-liq (Cou Test)
10ml
Legipid Kit 10 (LGP-10)
Measurement procedure:
1
Perform the measuring procedure according to the Biótica instructions.
2
Eluate filtered particles by shaking for 02:00 minute(s).
3
Tap TEST to start a 02:00 minute(s) countdown.
4
Tap TEST to start a 15:00 minute(s) countdown.
5
Tap TEST to start a 05:00 minute(s) countdown.
6
Tap TEST to start a 03:00 minute(s) countdown.
7
Tap TEST to start a 10:00 minute(s) countdown.
8
Tap TEST to start a 03:00 minute(s) countdown.
9
Tap TEST to start a 03:00 minute(s) countdown.
10
Tap TEST to start a 03:00 minute(s) countdown.
11
Tap TEST to start a 03:00 minute(s) countdown.
12
Tap TEST to start a 02:00 minute(s) countdown.
13
Tap TEST to start a 05:00 minute(s) countdown.
14
Insert the filled 1 ml LG-CB cuvette into the PrimeLab.
15
Put on the lightshield.
16
Start ZERO measurement.
17
Insert the filled 1 ml LG-CB cuvette into the PrimeLab.
18
Put on the lightshield.
19
Tap TEST to perform the measurement.
20
The measured value is immediately displayed.
Notes:
Result is displayed as “cfu/l” which is related to filtration of 1 litre of your sample.
Once reagents are received, kit MUST be stored b2 °C and +8 °C, preferrably at +4
°C.
Expiry date of the reagents is 3 months from production date on.
T111/204
T111/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...