(38)
pH-value MR
6.50 - 8.40
(pH)
Tablet
Internal Name: 38-pH-MR-tab
10ml
Phenol Red Photometer (TbsPph)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 1 Phenol Red Photometer tablet(s) to the test water in the cuvette.
8
Crush the tablet with a clean stirring rod.
9
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
10
Screw the lid back on the cuvette.
11
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
12
Tap TEST to perform the measurement.
13
The measured value is immediately displayed.
Notes:
PH value out of the range of pH 6.5 to 8.4 can lead to incorrect test results. If you are not sure,
a control measurement by pH electrode is recommended.
An alkalinity level > 40 mg/l CaCO₃ is required to prevent inaccurate readings.
Depending on the salt content of the water sample, the measurement result must be manually
corrected according to the following scheme: 1 molar = -0.21 pH; 2 molar = -0.26 pH; 3 molar =
-0.29 pH with: 1 mol of salt (NaCl) = 5.8 % = 58.4 g/l.
T151/204
T151/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...