(154)
Phosphorus-total HR
0.0 - 52.0
mg/l (P)
Powder Pack
+ Tablet
Internal Name: 154-PsphrTotHR-tab
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
Phosphate HR N°1
Photometer (PPHPPHR1)
Phosphate HR N°2
Photometer (TbsPPPHR2)
Measurement procedure:
1
Fill 5 ml sample water in a fresh, clean Phosphorus HR cuvette.
2
Add 2 x 0.05 ml PL Phosphorus 2 to the test water in the cuvette.
3
Screw the lid back on the cuvette IMMEDIATELY.
4
Shake the cell vigorously for 00:20 minute(s).
5
Tap TEST to start a 00:20 minute(s) countdown.
6
Place cuvettes for 30 minutes at 150 °C in the preheated thermoreactor.
7
Tap TEST to start a 30:00 minute(s) countdown.
8
CAUTION: Cuvettes are hot!
9
Remove the cuvettes from the thermoreactor.
10
Let the cuvettes cool down to at least 60 °C.
11
Add 10 drop(s) of PL Phosphorus HR1 into the cuvette.
12
Screw the lid back on the cuvette.
13
Gently swirl the cuvette to mix the liquid with the reagent.
14
Place the 16 mm adapter in the PrimeLab.
15
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
16
Start ZERO measurement.
17
Remove the cuvette again.
18
Unscrew the lid from the cuvette.
19
Add 1 Phosphate HR 1 powder pillow to the sample water in the cuvette.
20
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
21
Add 1 Phosphate HR 2 tablet(s) to the test water in the cuvette.
22
Crush the tablet with a clean stirring rod.
23
Screw the lid back on the cuvette.
24
Gently swirl the cuvette to mix the liquid well.
25
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
26
Tap TEST to perform the measurement.
T170/204
T170/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...