(150)
Urea HR
0.2 - 5.0
mg/l ((NH₂)₂CO)
Powder Pack
+ Liquid
Internal Name: 150-UreaHR-tab-liq
10ml
Ammonia N°1 Photometer (PPHAM1)
Ammonia N°2 Photometer (PPPAM2)
30ml PL Urea N°1 (PL30Urea1)
10ml PL Urea N°1 (PL10Urea2)
Measurement procedure:
1
Fill 5 ml distilled water into a clean 24 mm cuvette.
2
Add 5 ml test water to the same cuvette.
3
Screw the lid back on the cuvette.
4
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
5
Start ZERO measurement.
6
Remove the cuvette again.
7
Unscrew the lid from the cuvette.
8
Add 2 drop(s) of PL Urea 1 into the cuvette.
9
Screw the lid back on the cuvette.
10
Gently swirl the cuvette to mix the liquid well.
11
Add 1 drop(s) of PL Urea 2 into the cuvette.
12
Screw the lid back on the cuvette.
13
Gently swirl the cuvette to mix the liquid well.
14
Tap TEST to start a 05:00 minute(s) countdown.
15
The cuvette does not need to be placed in the device during this time.
16
Add 1 Ammonia N°1 Photometer powder pillow to the sample water in the cuvette.
17
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
18
Add 1 Ammonia N°2 Photometer powder pillow to the sample water in the cuvette.
19
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
20
Screw the lid back on the cuvette.
21
Gently swirl the cuvette to mix the liquid well.
22
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
23
Tap TEST to start a 10:00 minute(s) countdown.
24
After the lapse of a 10:00 minute(s) countdown the measured value is displayed.
Notes:
T200/204
T200/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...