(21)
DEHA
20 - 1000
µg/l (DEHA)
Liquid
Internal Name: 21-DEHA-liq
10ml
PL Oxygen Scavenger 1 (30 ml) (PL30OxyScav1)
PL Oxygen Scavenger 2 (65 ml) (PL65OxyScav2)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 6 drop(s) of PL Oxygen Scavenger 1 into the cuvette.
8
Gently swirl the cuvette to mix the liquid well.
9
Add 25 drop(s) of PL Oxygen Scavenger 2 into the cuvette.
10
Screw the lid back on the cuvette.
11
Gently swirl the cuvette to mix the liquid well.
12
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
13
Tap TEST to start a 10:00 minute(s) countdown.
14
After the lapse of a 10:00 minute(s) countdown the measured value is displayed.
Notes:
Ferrous sample water will interfere with this test and can influence the readings. To determine
the ferrous iron concentration for correction purposes, repeat the test without adding PL Oxygen
Scavenger N°1. If the result is above 0.05 mg/l, subtract this value from the DEHA result.
During the reaction time, ensure the sample is kept in the dark.
Shake the liquid reagent thoroughly before adding the liquid to the vial.
T71/204
T71/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...