(63)
Bromine
(in presence of chlorine)
0.00 - 9.00
mg/l (fBr₂)
Liquid
+ Powder
Internal Name: 63-Bromine-liq
10ml
30ml PL DPD 1 A (PL30DPD1A)
30ml PL DPD 1 B (PL30DPD1B)
PL DPD Nitrite Powder (PLpow20DPDNitr)
30ml PL DPD Glycine (PL30DPDGlycine)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 3 drop(s) of PL DPD Glycine into the cuvette.
8
Gently swirl the cuvette to mix the liquid well.
9
Add 3 drops of PL DPD 1 A and PL DPD 1 B into a second clean 24 mm cuvette.
10
Fill the cuvette with the treated sample of the first cuvette.
11
Screw the lid back on the cuvette.
12
Gently swirl the cuvette to mix the liquid well.
13
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
14
Tap TEST to perform the measurement.
15
The measured value for tBr₂ (total bromine) is immediately displayed.
16
If you require determination of ´combined´ and/or ´free´ bromine, please proceed with the
following steps.
17
Fill 10 ml test water into a clean 24 mm cuvette.
18
Add 1 x 0.05 ml PL DPD Nitrite scoop(s) powder to the sample water in the cuvette.
19
Screw the lid back on the cuvette.
20
Gently swirl the cuvette to mix the liquid well.
21
Add 3 drops of PL DPD 1 A and PL DPD 1 B into a second clean 24 mm cuvette.
22
Fill the cuvette with 10 ml of the treated sample water from the first cuvette.
23
Screw the lid back on the cuvette.
24
Gently swirl the cuvette to mix the liquid well.
25
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
26
Tap TEST to perform the measurement.
T17/204
T17/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...