(170)
Transmission
0.0 - 100.0
% (Trnsm)
Internal Name: 170-Transmission
10ml
Measurement procedure:
1
Select the desired wavelength.
2
Fill 10 ml test water into a clean 24 mm cuvette.
3
Screw the lid back on the cuvette.
4
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
5
Start ZERO measurement.
6
Remove the cuvette again.
7
Treat the water sample according to the chosen procedure.
8
Fill 10 ml of the treated water sample into a clean 24 mm cuvette.
9
Screw the lid back on the cuvette.
10
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
11
Tap TEST to perform the measurement.
12
The measured value is immediately displayed.
Notes:
With this method you can create your own parameters, use reagents from other manufacturers
and/or perform measurements with the PrimeLab that are not included in the offered IDs/
parameters.This requires that you familiarize yourself with the colorimetry of the water sample
AFTER addition of the reagent that you want to use. Select the wavelength of your sample after
addition of the reagent to be used by selecting the closest colour match (see also
www.primelab.org).At the end of measurement you will receive a value for the "Transmission".
"Transmission" in % means how much light reaches the sensor (in %), compared to ZERO
measurement (T = 100%). After adding a colouring reagent, transmission will decrease.Simply
measure several water samples with different concentrations of the parameter of interest on one
wavelength, to record your own values using the determined transmission results.
T193/204
T193/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...