(128)
Bromine
0.00 - 4.50
mg/l (Br₂)
Powder Pack
Internal Name: 128-Bromine-pp
10ml
DPD N°1 Photometer (PPPD1)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 1 DPD N°1 Photometer powder pillow to the sample water in the cuvette.
8
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
9
Screw the lid back on the cuvette.
10
Gently swirl the cuvette to mix the liquid well.
11
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
12
Tap TEST to start a 03:00 minute(s) countdown.
13
After the lapse of a 03:00 minute(s) countdown the measured value is displayed.
Notes:
Turbidity caused by high concentration of calcium ions will affect the measurement result. To
prevent that, please use DPD HC (High Calcium) reagents.
Make sure no bromine escapes while preparing the sample. The measurement must be
performed directly after sampling.
If the sample water contains more than 40 mg/l bromine, a measured value of 0 mg/l can be
displayed. In that case, a dilution is recommended.
The DPD reagent buffers the pH value of the sample water in the range of 6.2 to 6.5, which is
essential for colour development. If the sample water is very alkaline or acidic, this must be
adjusted to a pH within the range of 6 to 7 by addition of 0.5 mol/l sulphuric acid or 1 mol/l
caustic soda, respectively, before the DPD reagent is added.
If the sample water contains further oxidizing agents, these will react like bromine and
contribute to the measurement result.
T19/204
T19/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...