(169)
Nitrate HR
1 - 100
mg/l (NO₃⁻)
Powder Pack
Internal Name: 169-Nitrate-HR-pp
10ml
Nitrate N°1 Photometer (PPHNitra1)
Nitrate N°2 Photometer (PPPNitra2)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette from the PrimeLab and set it aside. It is no longer required for this test.
6
Fill 20 ml sample water in a test tube.
7
Add a Nirate N°1 Photometer powder pillow to the test tube.
8
Then add a Nitrate N°2 Photometer powder pillow to the test tube.
9
Screw the lid back on the tube.
10
Shake the tube heavily for 00:15 minute(s).
11
Tap TEST to start a 10:00 minute(s) countdown.
12
Wait until the 10:00 minute(s) countdown ran out.
13
Remove 10 ml from the tube by using a syringe.
14
Empty the 10 ml from the previous step in a clean 24 mm cuvette.
15
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
16
Tap TEST to perform the measurement.
17
The measured value is immediately displayed.
Notes:
In presence of nitrites, measurements of excessively high nitrate levels are possible. For
correction, subtract the nitrite-nitrogen (N) from the content of the nitrate-nitrogen (N). For this,
the measurement results of the nitrate or nitrite measurement must be calculated in nitrogen
(N).
Too short or too weak shaking of the sample may result in lower nitrate levels.
If an uneven color distribution in the shaker is observed within the waiting time, it should be
reversed again. The current countdown remains unaffected.
T126/204
T126/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...