(163)
Dissolved Oxygen
0.0 - 10.0
mg/l (O₂)
Liquid
Internal Name: 163-Dis.Oxygen
10ml
30ml PL Dissolved Oxygen N°1 (PL30DO1)
30ml PL Dissolved Oxygen N°2 (PL30DO2)
30ml PL Dissolved Oxygen N°3 (PL30DO3)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Empty the cuvette.
8
Clean the cuvette.
9
Clean the lid as well.
10
Fill a 50 ml glass bottle brimful with the water to be measured.
11
Place the stopper on the glass bottle. Caution, excess water runs out!
12
Remove the stopper again.
13
Add 10 drop(s) of PL DissOx 1 to the glass bottle.
14
Put the stopper back on.
15
Swivel/shake glass bottle for 01:00 minute(s)
16
Tap TEST to start a 01:00 minute(s) countdown.
17
Remove the stopper again.
18
Add 10 drop(s) of PL DissOx 2 to the glass bottle.
19
Put the stopper back on.
20
Swivel/shake glass bottle for 01:00 minute(s)
21
Tap TEST to start a 01:00 minute(s) countdown.
22
Remove the stopper again.
23
Add 10 drop(s) of PL DissOx 3 to the glass bottle.
24
Put the stopper back on.
25
Swivel/shake glass bottle for 01:00 minute(s)
26
Tap TEST to start a 01:00 minute(s) countdown.
T72/204
T72/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...