(59)
Turbidity
20 - 1000
FAU (Turb)
Internal Name: 59-Turbidity
10ml
Measurement procedure:
1
Fill 10 ml distilled water in a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Empty the cuvette.
8
Mix the sample thoroughly.
9
Rinse the cuvette several times with the sample water.
10
Then fill 10 ml of the sample water in the cuvette.
11
Screw the lid back on the cuvette.
12
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
13
Tap TEST to perform the measurement.
14
The measured value is immediately displayed.
Notes:
The test result can be converted into the following unit(s): FTU (same as FAU)
FAU stands for Formazin Attenuation Units, different from the NTU (nephelometric) method.
The measurement should be conducted immediately after sampling.
Air bubbles will influence the measurement result.
Tinted water samples influence the result. In this case do not use distilled water (step 1) but
rather filtered sample water for the ZERO adjustment.
T194/204
T194/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...