(110)
Phosphonate
0.0 - 20.0
mg/l (PO₄³⁻)
Tablet
Internal Name: 110-Phosphon-tab
10ml
Oxidising OP Photometer
(TbsHOXOP)
OP-A Photometer (TbsPOPA)
OP-B Photometer (TbsPOPB)
OP-AX Photometer (TbsHOPAX)
Measurement procedure:
1
Fill 8 ml of distilled water into a clean 24 mm cuvette.
2
Add exactly 2 ml sample water to the same cuvette.
3
Screw the lid back on the cuvette.
4
Swivel the cuvette back and forth for 5 times.
5
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
6
Start ZERO measurement.
7
Remove the cuvette again.
8
Unscrew the lid from the cuvette.
9
Add 1 OrgaPhos-OX tablet(s) to the test water in the cuvette.
10
Crush the tablet with a clean stirring rod.
11
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
12
Screw the lid back on the cuvette.
13
Gently swirl the cuvette to mix the liquid well.
14
Tap TEST to start a 05:00 minute(s) countdown.
15
The cuvette does not need to be placed in the device during this time.
16
Remove the cuvette again.
17
Unscrew the lid from the cuvette.
18
Add 1 OrgaPhos No.1 tablet(s) to the test water in the cuvette.
19
Crush the tablet with a clean stirring rod.
20
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
21
Screw the lid back on the cuvette.
22
Gently swirl the cuvette to mix the liquid well.
23
Tap TEST to start a 02:00 minute(s) countdown.
24
The cuvette does not need to be placed in the device during this time.
25
Remove the cuvette again.
26
Unscrew the lid from the cuvette.
T166/204
T166/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...