Permanganate Time Test
0 - 100
%T (PTT)
Liquid
Internal Name: 159-PTT-tab
10ml
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
Potassium Permanganate Solution
(PL10PTT)
Measurement procedure:
1
Set the refrigerator to 15 °C using the thermostat and according to the instructions.
2
Fill 10 ml test water into a clean 24 mm cuvette.
3
Screw the lid back on the cuvette.
4
Place the cuvette in the refrigerator for 20 minute(s).
5
Place the sealed "Methanol ZERO" cuvette in the PrimeLab.
6
Start ZERO measurement.
7
Remove the cuvette from the PrimeLab and set it aside. It is no longer required for this test.
8
Remove the cuvette from the refrigerator.
9
Add exactly 35 µl of Potassium Permanganate Solution to the sample water into the same
cuvette. Use a 10-100 µl pipette.
10
Screw the lid back on the cuvette.
11
Place the cuvette in the refrigerator for 10 minute(s).
12
Tap TEST to start a 10:00 minute(s) countdown.
13
Remove the cuvette from the refrigerator.
14
Wipe the condensation water from the cuvette with a dry cloth.
15
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
16
Tap TEST to perform the measurement.
Notes:
After adding the PTT-liquid, sample is highly sensitive to light, air and temperature. Do NOT
open the vial after PTT-liquid has been added and lid got screwed on and keep it at constant
temperature of 15 °C.
Interferences: Turbid and/or colored water samples (before adding PTT liquid).
T150/204
T150/204
Содержание Primelab 2.0
Страница 12: ...11 Icons Bluetooth WiFi Display General Settings Sound Device Information Cloud Sampling Points PrimeLab2 0...
Страница 13: ...Favourites Calculator Main Menu Battery Search General Create New Account Edit Return Icons 12 PrimeLab2 0...
Страница 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Страница 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Страница 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Страница 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Страница 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Страница 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Страница 179: ...Cobalt levels above 0 5 mg l will interfere with this test and can influence the reading high reading T124 204 T124 204...
Страница 237: ...The precision of the test results depends upon the precision of the diluting procedure T182 204 T182 204...
Страница 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Страница 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Страница 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Страница 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Страница 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...