AUTOMATIC WATER SUPPLY
Your sterilizer can be connected to an external water supply system and to a continuous drain line. Once
connected, the clean water tank will be automatically filled when empty, and the used water will drain con-
tinuously. Thus, neither manual filling nor draining are required. The water supply system must deliver de-
mineralized water meeting the requirements as listed in Appendix 7.
4.5.1
Connection to the water supply and
to the drain
Remove the two caps installed on the elbow fittings
on the rear side of the sterilizer. Connect the water
supply to the D2 port using a 6mm ext/4 mm int. dia.
tube (not included). The water pressure must be be-
tween 2 and 8.6 bar, max. flow 0.5 l/min.
Connect the drain tube to the D4 port using an 8mm
ext/6mm int. dia. plastic tube: a 3 meter piece of tube
is included in the package.
4.5.2
Demineralizer (optional)
W&H’s Multidem is the water filtration system de-
signed to provide high quality demineralized water
for the sterilization process and for all the other
needs of the dental practice. Contact your dealer for
more information.
4.5.3 Drain tube mounting kit with fittings
(optional)
This kit (see image to the left) allows to connect the
sterilizer drain tube to an existing drain pipe in a fast
and easy way. See Appendix 9.