51
11.5 CLEANING THE EXTERNAL STERILIZER SURFACES
•
Disconnect the mains cable.
•
Never use disinfectants or abrasive products.
•
Clean all external covers with a damp cloth and a mild detergent solution.
•
Do not use excessive amounts of water to wash the sterilizer as this may
damage the electrical components and safety mechanisms.
•
Take care not to scratch the plastic film in front of the touch-screen; avoid clean-
ing it with detergents or pointed objects.
11.6 REPLACING THE BACTERIOLOGICAL FILTER
•
Open the service door.
•
Unscrew the bacteriological filter by hand (counter-clockwise).
•
Screw on the new filter (clockwise) and tighten it snug.
ATTENTION!
Disposal of used consumables shall be done in
compliance with local laws and rules
11.7 REPLACING THE DUST FILTER
•
Pull out the filter from the bottom/front of the ste-
rilizer.
•
Detach the used filter from the handle.
•
Attach the new filter to the handle.
•
Slide the filter back into its position.
ATTENTION!
Disposal of used consumables shall be
done in compliance with local laws and
rules