46
9.
ALARMS
Lisa sterilizers constantly monitor a variety of parameters during the sterilization cycle to assure a proper
sterilization of the load. In case any of the parameters are off, potentially affecting the sterilization efficien-
cy, the cycle is immediately interrupted and an alarm message is displayed.
Once the sterilizer generates an alarm message, a two-
minute reset phase starts to drain the pneumatic circuits
and depressurizes the chamber. Do not switch off or un-
plug the sterilizer during this reset phase!
At the end of the reset phase, the door can be unlocked by
pressing the “Unlock door” icon. The screen will display the
type of alarm that was generated and indicate the condition
of the sterilization load. Refer to § 10 for alarm-code troub-
leshooting.
If an alarm message is generated before the sterilization/plateau phase (PR) of the
sterilization cycle is completed, the following message will be displayed:
!! Load not sterile !!
If an alarm message is generated after the completion of the sterilization/plateau
phase of the sterilization cycle (e.g., during the drying phase), the following message
will be displayed:
!! Sterile conditions achieved, drying interrupted !!
Both messages are printed or saved on the memory card.