48
Code
Description
Action
Vacuum pump
A310
During a vacuum phase of the cycle, a vacuum of
–0.20 bar or lower could not be achieved.
Clean, check or replace the door seal (§11.2) and the
face side of the chamber; repeat the cycle.
If the problem persists
service.
A320
During a vacuum phase of the cycle, a vacuum of
–0.50 bar or lower could not be achieved.
A330
Vacuum pump fault
service
A350
Vacuum pump fault
service
A380
Leak between the reading points P1 and P2 of the
Vacuum Test
Clean, check or replace the door seal (see § 11.2)
and the face side of the chamber; repeat the cycle. If
the problem persists
service.
Note: Ensure that the sterilizer chamber is cold and
dry when initiating Vacuum Tests!
A390
Leak between the reading points P2 and P3 of the
Vacuum Test
Water pump
A400
Water pump fault
service
A410
Water conductivity meter fault
service
A420
Flow meter fault
service
Door locking
A520
The chamber door locking mechanism blocked
during opening.
service
A530
The chamber door locking mechanism blocked
during closing.
service
Electric valves
A601…8 Electric valve fault
service
Condenser
A720
Fan fault
service
Water tanks
A810
Water level sensor fault
service
Manual stop
A990
The “Manual stop” icon was pressed
Wait until the end of the reset phase (see § 7.2).