10.
ALARM CODE TABLE
Refer to the table below for troubleshooting assistance in case the sterilizer generates an alarm code.
Code
Description
Action
Power supply
A010
Power failure or significant drop in voltage occurred
during the cycle.
The load cannot be considered sterile. The cycle
must be repeated. Check building power supply to
the sterilizer.
Sterilization chamber
A100
The time it took to reach the sterilization plateau is
too long (overload, leaks, etc.).
Observe the maximum load weight limits, clean,
check or replace the door seal and the face side of
the chamber and repeat the cycle.
If the problem persists
service.
A130
During the sterilization phase of the cycle, the pres-
sure measured inside the sterilization chamber is
above the maximum threshold.
Repeat the cycle.
If the problem persists
service.
A140
During the sterilization phase of the cycle, the pres-
sure measured inside the sterilization chamber is
below the minimum threshold.
Clean, check or replace the door seal and the face
side of the chamber and repeat the cycle.
If the problem persists
service.
A150
During the sterilization phase of the cycle, the tem-
perature of the steam is below the minimum thre-
shold.
Clean, check or replace the door seal and the face
side of the chamber and repeat the cycle.
If the problem persists
service.
A160
During the sterilization phase of the cycle, the tem-
perature of the steam is above the maximum thre-
shold.
Repeat the cycle.
If the problem persists
service.
A170
The temperature sensor of the chamber heating
element is broken or disconnected.
service
A180
The internal temperature sensor of the chamber is
broken or disconnected.
service
A190
Air detector alarm
Clean, check or replace the door seal and the face
side of the chamber and repeat the cycle.
If the problem persists
service.
Steam generator
A230
The temperature sensor of the steam generator is
broken or disconnected.
service
A240
Fault of the steam generator heater.
service
A250
The internal temperature sensor of the air detector
system is broken or disconnected.
service
Содержание Lisa
Страница 1: ...Instruction for use UM009ENG LISA FULLY AUTOMATIC 201 10 Rev 4...
Страница 2: ......
Страница 77: ......