Instructions for Use
5
Unit Installation and Operating Environment
DANGER
WARNING
The Co-mix unit is designed for indoor laboratory use only.
The acceptable operating temperature range is 10ºC to 38ºC,
with a relative humidity of 20% to 85% non-condensing, at a
maximum altitude of 2000m above sea level.
If the unit is stored in conditions outside of these ranges, it
must be left to stand unpowered until it has acclimatised to
within these environmental limits before being powered.
DANGER
Use only the AC mains power cord provided with the unit or as
specified in section 11.
The unit must be connected to a suitably earthed mains
supply, with appropriate earth-leakage and over-current
protection.
WARNING
Always ensure that the mains power connector is securely
inserted into the rear of the unit, and any excess power cord
does not pose a potential trip or pull hazard.
DANGER
Do not operate the unit in any area which is, or has been, or is
thought to have been exposed to explosive or flammable
gases, vapours or liquids.
WARNING
The unit must be installed and operated on a solid, stable,
vibration-free and level working surface; ensuring that the
ventilation slots on the underside of the unit are not blocked.
CAUTION
The unit can create strong vibrations at high speeds which
may cause objects positioned near to it to move.
Do not place easily movable objects near the unit; or ensure
that such objects are appropriately secured.